1mA – 0.99A Custom AC/DC High precision current shunt

  • AC/DC differences from 1ppm
  • Shunt values for measurements between 1mA to 100A
  • From DC to 30kHz
  • Low inductance & capacitance flat frequency response
  • Simplifies precision calibration of calibrators and current sources
  • Can be used with a precision multimeter or an AC measurement standard
SKU: 1mA - 0.99A Category:
Brands:
Brands

Description

1mA – 0.99A Custom AC/DC High precision current shunt

Key Features

  • AC/DC differences from 1ppm
  • Shunt values for measurements between 1mA to 100A
  • From DC to 30kHz
  • Low inductance & capacitance flat frequency response
  • Simplifies precision calibration of calibrators and current sources
  • Can be used with a precision multimeter or an AC measurement standard

Description

Transmille offers a set of current shunts which enable measurement of current from 1mA  to 100A. Each shunt develops 0.7V at nominal full scale, enabling a wide range of multimeters to be used without loss of resolution due to changing range.

 

 

 

High Performance Design

Transmille’s AC/DC current shunts provide excellent DC value stability, but also a wide range in frequency allowing measurements from DC up to and beyond 30kHz. The coaxial design of these shunts reduces inductive and capacitive effects, while an open design improves heat dissipation to minimize power co-efficients. The coaxial design offers superior accuracy and low electromagnet influences, ensuring a linear and predictable frequency response. Each shunt has a type N output connector to ensure low noise when measuring at high frequencies, providing easy connections from the shunt to the measuring device.

The set of AC/DC current shunts has been designed to enable fast and efficient calibration of multiproduct calibrators and current sources under a laboratory environment. Using the AC/DC current shunts in conjunction with an 8000 series multimeter, the 3000A, 4000 and 9000A multiproduct calibrators can be efficiently maintained internally, as well as supporting other manufacturers’ products.

  • AC/DC differences from 1PPM
  • Superior accuracy compared to Wilkinson standard AC/DC resistors, minimizing the use of correction factors for frequency response
  • Shunt values for measurements between 1mA to 100A
  • From DC to 30kHz
  • Low inductance & capacitance flat frequency response
  • Simplifies precision calibration of calibrators and current sources
  • Can be used with a precision multimeter, or an AC measurement standard

Transmille offers a total of 14 high-precision shunts, each with an output value of 0.7V against nominal input. These shunts come in various input values, from 1mA  to 100A*.

Current shunts can be supplied with certified AC/DC corrections from NPL for traceability directly to international standards.

An optional connections kit is also available to enable fast connections to typical test equipment.

*Please note: custom values are available upon request

1A AC/DC Current Shunt Certificate 20A AC/DC Current Shunt Certificate 100mA AC/DC Current Shunt Certificate
SHUNT

NOMINAL VALUE (OHMS)

1x 1mA High Precision coaxial shunt 714
1x 10mA High Precision coaxial shunt 71.4
1x 20mA High Precision coaxial shunt 35.7
1x 50mA High Precision coaxial shunt 14.28
1x 100mA High Precision coaxial shunt 7.14
1x 200mA High Precision coaxial shunt 3.57
1x 500mA High Precision coaxial shunt 1.42
1x 1A High Precision coaxial shunt 0.714
1x 2A High Precision coaxial shunt 0.357
1x 5A High Precision coaxial shunt 0.142
1x 10A High Precision coaxial shunt 0.0714
1x 20A High Precision coaxial shunt 0.01428
1x 50A High Precision coaxial shunt 0.00357
1x 100A High Precision coaxial shunt 0.00714