Dual-Channel Precision Source Meter S3022F

The dual-channel high-precision source meter provides higher output capability and higher accuracy, and the rich LCD screen operation interface greatly improves the test efficiency.

pdf Semight_S3022F_Specification_V1.5_EN

SKU: S3022F Category:
Brands:
Brands

Description

Dual-Channel Precision Source Meter S3022F

High range

Range:±200 V、±3 A(direct)、±10A(pulse)

High resolution

The minimum measurement resolution can reach 100fa / 100nv

High sampling rate

Up to 1M ADC sampling rate

Threshold trigger

Hardware high-speed IO, which can realize threshold triggering and realize efficient interaction between output measured value and user system

Functions and advantages

Integration of five functions

Integration of five functions
Current source
Ammeter
Voltmeter
Electronic load

1. The first and third quadrants are the source: the actual polarity of the output V/I follows the source setting;

2. The second and fourth quadrants are the load: CC and CV cooperate, when the load is used, the load setting polarity is opposite to the source polarity;

Can test various equipment

Capture more measurement data

♦ 6.5-bit digital resolution: accuracy equivalent to6-and-a-half digitsA multimeter;
♦ 100 fa / 100 NV resolution: excellent sensitivity for setting and measuring;
♦ 1m points / second: provide high-speed measurement, and can quickly set / digitize the rate to any waveform generator / list scan;

Rich scanning function

Additional information

Weight 7.5 kg
Dimensions 44.9 × 44.1 × 11.25 cm

Voltage source index

Voltage setting accuracy Range Set resolution Accuracy (1 year) ± (% reading + offset) Typical noise (effective value)0.1 Hz-10Hz
±200 V 1mV 0.02%+40 mV 600 μV
±20 V 100 μV 0.02%+5 mV 100 μV
±6 V 10 μV 0.02%+500 μV 4 μV
±200 mV 1 uV 0.02%+200 μV 2 μV
Temperature coefficient ±(0.15 × Accuracy)/°C (0℃-18℃,28℃-50℃)
Maximum output power of single channel 30W: ±20V@1.5A;30W: ±200 V@0.15A;18W: ±6 V@3A
Set time <800uS (typical value)
Overshoot <±0.1% (typical value, normal, step is 10% to 90% of the range, full scale point, resistive load test)
Noise 10Hz-20MHz 6V voltage source, 3A resistance load,<3mV RMS

Current source index

Current setting accuracy Range Set resolution Accuracy (1 year) ± (% reading + offset) Typical noise (effective value)0.1 Hz-10Hz
±10 A1 50 μA 0.4% + 40 mA NA
±3 A 20 μA 0.05% + 5 mA 10 μA
±1.5A 5 μA 0.02% + 500 μA 3 μA
±150 mA 500 nA 0.02% + 25 μA 800 nA
±15 mA 50 nA 0.02% + 2.5 μA 100 nA
±1.5 mA 5 nA 0.02% + 150 nA 20 nA
±150 μA 500 pA 0.02% + 25 nA 200 pA
±15 μA 50 pA 0.02% + 3 nA 75 pA
±1.5 μA 5 pA 0.03% + 600 pA 50 pA
±150 nA 500 fA 0.05% + 300 pA 10 pA
Temperature coefficient ±(0.15 × Accuracy index)/°C (0℃-18℃,28℃-50℃)
Maximum output power of single channel 30W: ±20V@1.5A;30W: ±200 V@0.15A;18W: ±6 V@3A
Set time <500us (typical value)
Overshoot <±0.1% (Typical values,Normal,Step is 10% to 90% of the range, full scale point, resistive load test)

1, 10A The range only supports pulse mode, and the accuracy is typical

Voltmeter index

Voltage measurement accuracy Range Display resolution Accuracy (1 year) ±(% Reading + offset)
±200 V 100 μV 0.02% + 40 mV
±20 V 10 uV 0.02% + 5 mV
±6 V 1 uV 0.02% + 500 uV
±200mV 100 nV 0.02% + 200 μV
Temperature coefficient ±(0.15 × Accuracy)/°C (0℃-18℃,28℃-50℃)

Ammeter index

Current measurement accuracy Range Display resolution Accuracy (1 year) ±(% Reading + offset)
±10 A1 10 μA 0.4% + 25 mA
±3 A 10 μA 0.05% + 5 mA
±1.5A 1 μA 0.02% + 500 μA
±150 mA 100 nA 0.02% + 25 μA
±15 mA 10 nA 0.02% +2 μA
±1.5 mA 1 nA 0.02% + 150 nA
±150 μA 100 pA 0.02% + 20 nA
±15 μA 10 pA 0.02% +3 nA
±1.5 μA2 1 pA 0.03% + 600 pA
±150 nA2 100fA 0.05% + 300 pA
Temperature coefficient ±(0.15 × Accuracy)/°C (0℃-18℃,28℃-50℃)

1, 10A range only supports pulse mode, and the accuracy is typical
2, For small current measurement, it is recommended to use three coaxial cables for connection Force hi is connected to the core wire, guard is connected to the inner shield layer, the outer shield layer is connected to the protective ground, force Lo is connected to the core wire, the inner shield layer is not connected, and the outer screen
The shielding layer shall be grounded and the rated insulation voltage of the coaxial line shall not be less than 250V.

Ohmmeter index (4-wire)

Resistance measurement accuracy Range Display resolution Default measurement current Typical accuracy (1 year)
±(% Reading + offset)
1 Ω 1 uΩ 1 A 0.09% + 0.5 mΩ
10 Ω 10 uΩ 100 mA 0.065% + 5 mΩ
100 Ω 100 uΩ 10 mA 0.065% + 50 mΩ
1 KΩ 1 mΩ 1 mA 0.055% + 500 mΩ
10 KΩ 10 mΩ 100 uA 0.065% + 5 Ω
100 KΩ 100 mΩ 10 uA 0.07% + 50 Ω
1 MΩ 1 Ω 1 uA 0.05% + 500 Ω
10 MΩ 10 Ω 0.1 uA 0.65% + 5KΩ
100 MΩ 100 Ω 0.05 uA 1.27% + 10 KΩ
Temperature coefficient ±(0.15 × Accuracy)/°C (0℃-18℃,28℃-50℃)
Manual current source resistance measurement (four wire) Overall error = measured voltage / current source set current = resistance reading x (gain error percentage of voltage source range + gain error percentage of ammeter range + current source range bias error / set current) + (voltage source range bias error / set current value)
Example: current source setting current = 1A voltage measuring range = 6V
Overall error=(0.02%+0.02%+500uA/1A)+(500uV/1A)
=0.09%+0.5mΩ

Pulse source index (4 lines)

Minimum programmable pulse width 100μs
Pulse width programming resolution 1μs
Pulse width programming accuracy ±10μs
Pulse width jitter 2μs
Pulse width definition As shown in the following figure, the time from 10% leading edge to 90% trailing edge

Pulse technical index Maximum current limit Maximum pulse width Maximum duty cycle
1 0.15A/200V DC,unlimited 100%
2 1.5A/20V DC,unlimited 100%
3 3A/6V DC,unlimited 100%
4 3A/20V 1mS 10%
5 10A/6V 1mS 10%

Rise time of pulse source (4 lines)

Output Range Typical rise time 1,3 Typical stabilization time 2,3 Test load
Voltage source 200 V 600 μS 1.5 mS No load
20 V 200 μS 360 μS No load
6 V 160 μS 300 μS No load
Current source 10 A 140 μS 320 μS The full load
3 A 120 μS 280 μS The full load
1.5 A 120 μS 280 μS The full load
150 mA 120 μS 280 μS The full load
15 mA 120 μS 280 μS The full load
1.5 mA 120 μS 280 μS The full load

1, The time required for the pulse front from 10% to 90%.
2, The time required for the pulse to reach 1% of the final value.
3,The current source test is based on the test at 6V voltage range, with 105% full scale clamp setting.

I-V Output capacity

Output setup time

Output Range Typical output setup time Test conditions:
Fast 1,2 Normal 1 Slow1
Voltage source 200V <1.3mS <1.5mS <2.5ms The time required to reach within 0.1% of the final value under open circuit load. In steps ranging from 10% to 90%.
20V <300μS <360μS <1ms
6V <150μS <250μS <1ms
200mV <200μS <250μS <1ms
Current source 3A <200μS <280μS <1.2ms The time required to reach within 0.1% of the final value (0.3% for the 3 a range) under short circuit conditions. The step is 10% to 90% of the range. 6V voltage range, 105% full scale clamp setting
1.5A <200μS <280μS <1.2ms
150mA <200μS <280μS <1.2ms
15mA <200μS <280μS <1.2ms
1.5mA <200μS <280μS <1.2ms
150μA <250uS <300uS <1.2ms
15μA <250uS <1.2mS <2ms
1.5uA <600uS <1.2mS <5mS
150nA <600uS <5mS <12mS

1, Output conversion rate:Fast,Normal,Slow。
2, Fast The mode output may have large overshoot under different range or load conditions. It is recommended to use slow mode for overshoot sensitive equipment.

Sampling rate and NPLC settings

Collocation method Configuration scope
NPLC 0.00005PLC ~ 10PLC
Sampling Rate 5sps ~ 1Msps

Derating of measurement accuracy(PLC<1)
Error: percentage of range increase

PLC Range
200mV 6V 20 to 200V 150nA to 1.5uA 15uA 150uA to 150mA 1.5A to 3A
0.1 0.02% 0.01% 0.01% 0.02% 0.01% 0.01% 0.01
0.01 0.30% 0.03% 0.02% 0.20% 0.04% 0.02% 0.02%
0.001 3.20% 0.40% 0.10% 2.50% 0.40% 0.03% 0.03%

Supplementary features

Sensing mode 2Line (2W)Or 4-wire (4W remote sensing) connection
Maximum sensing lead resistance 1 kΩ(accuracy rating)
Internal line pressure drop of line 2 <60mV/A
Maximum voltage between remote sensing output terminal and sensing terminal 2V
Maximum output voltage of output connector >Full scale 105%(200V range>202V)
DC floating ground voltage ±250 V DC
SWEEP scanning The scanning interval is configurable from 20us to 16S, and the maximum scanning point is 64K
Auto Range Support. It is recommended to switch off the output before switching the range for overshoot sensitive equipment
Delay measurement(SOURCE DELAY) Support. It is recommended that the user set the appropriate source delay to obtain more accurate measurement values
Over temperature protection When the internal temperature is detected to be too high, the output will be turned off, and the operation will be resumed when the temperature returns below 65 ℃
Overvoltage protection(OVP) When the output exceeds the OVP set value, turn off the output. After the equipment is reset, it can be operated further. The accuracy:±(1%Setting+500mV)
Other output abnormal protection Power off and restart to recover operation or hardware damage

Communication port

Ethernet 100BASE-T / 10BASE-T
USB USB 2.0 HOST(front)
USB 2.0 DEVICE(after)
 

Number I/O DB9

Absolute maximum input voltage:5.25

Absolute minimum input voltage:-0.25V

Minimum logic high level:2.1V

Highest logic low level:0.7V,

Maximum logic output current:2mA

Maximum suction current:-50mA

 

Pin5 land
Pin6 IO1, digital I / O, synchronous signal input port
Pin7 Io2, digital I / O, synchronous signal output port
Pin8 IO3, channel 2 digital I / O, synchronous signal output port (dual channel synchronous mode), synchronous signal
Input port (single channel synchronous mode)
Pin9 IO4, channel 2 digital I / O, single channel synchronous mode, synchronous signal output port

Environmental indicators

Environment Use in indoor facilities
Work 0 ° C to + 50 ° C, 30% to 70% relative humidity non condensing
Store -30 ° C to 70 ° C, 10% to 90% relative humidity non condensing
Altitude Working height: 0 m to 2000 m, storage: 0 m to 4600 M
Power Supply 90 V to 264 V, 47 Hz to 63 Hz, 250 VA Max
Preheat 1 hour
Size 429 × 441× 112.25mm(Including foot pad, handle and knob)
Weight Net weight 7.5 kg

front panel

Display 4.3 inch TFT LCD screen, capacitive touch mode, resolution 480 * 272
Fixed function key Trigger,Home,Cancel,Power on button, channel switch, rotary button
Non fixed function key LCD Mapping function keys
Interface USB Host interface, output interface, chassis ground interface

Rear panel

Interface Single channel precision power / measurement unit, pulse source Ethernet interface, DB9 interface, USB device interface, AC switch input socket, fan,
Chassis ground screw