4-Channel PXIe Precision Source Meter S0342C

4-channel high-precision PXIe source measure module, supports most of pxie chassis, It is convenient to integrate and extend to large integrated test system.

pdf Semight_S0342C_Specification_V1.2_EN

SKU: S0342C Category:
Brands:
Brands

Description

4-Channel PXIe Precision Source Meter S0342C

Easy to expand

Single channel standard pxie precision power supply, easy to expand

Large range

Range: ± 30 V, ± 500 MA (DC)

High resolution

The minimum measurement resolution can reach 100pA/100uV

High sampling rate

Up to 500K ADC sampling rate

Functions and advantage

(5) function in one body

Voltage source
Current source
Ammeter
Voltmeter
Electronic load

1、 Three quadrants as the source: the actual polarity of output V / I follows the source setting;
2、 The four quadrants are loads: CC and CV cooperate. When the load is used, the polarity of the load setting is opposite to the source polarity;
Can test various equipment

Capture more measurement data

♦ 6-digit digital resolution: accuracy equivalent to 6-digit digital multimeter;
♦ 100Pa / 100uv resolution: excellent measurement sensitivity;
♦ 500K points / second sampling rate: provide high-speed measurement, and can quickly set / digitize the rate to any waveform generator (list scan);

Rich scanning function

Voltage source index

Voltage setting accuracy Range Set resolution Accuracy (1 year) ± (% reading + offset) Typical noise (effective value)0.1 Hz-10Hz
±30V 500 uV 0.03%+4 mV 1000 μV
±6V 100 uV 0.03%+1 mV 100 μV
Temperature coefficient ±(0.15 × Accuracy)/°C (0℃-18℃,28℃-50℃)
Channel 1 CH0 reach CH3
Maximum output power of single channel 3W: ±30V@0.1A,±6V@0.5A
Set time <200uS (Typical value)
Overshoot <±0.1% (Typical value, normal, step is 10% to 90% of the range, full scale point, resistive load test)
Noise 10Hz-20MHz 6V voltage source, 0.5A resistive load,<3mV RMS

1,All channel outputs are electrically isolated from the earth, but all channel outputs are grounded(LO)

Current source index

Current setting accuracy Range Set resolution Accuracy (1 year) ± (% reading + offset) Typical noise (effective value)0.1 Hz-10Hz
±500 mA1 10 uA 0.05% + 100 μA 10 uA
±100 mA 2 uA 0.05% + 10 μA 1 uA
±10 mA 200 nA 0.05% + 5 μA 100 nA
±1 mA 20 nA 0.05% + 500 nA 10 nA
±100 μA 2 nA 0.05% + 50 nA 1 nA
±10 μA 200 pA 0.05% + 20 nA 150 pA
Temperature coefficient ±(0.15 × Accuracy)/°C (0℃-18℃,28℃-50℃)
Channel 2 CH0 reach CH3
Maximum output power of single channel 3W: ±30V@0.1A,±6V@0.5A
Set time <300us (Typical value)
Overshoot <±0.1% (Typical value, normal, step is 10% to 90% of the range, full scale point, resistive load test)

1,500mA range only supports 6V voltage range
2,All channel outputs are electrically isolated from the earth, but all channel outputs are grounded(LO)

Voltmeter index

Voltmeter accuracy Range Measurement resolution Accuracy (1 year) ± (% reading + offset)
±30V 1mV 0.03%+4 mV
±6V 100 uV 0.03%+1 mV
Temperature coefficient ±(0.15 × Accuracy)/°C (0℃-18℃,28℃-50℃)

Ammeter index

Accuracy of ammeter Range Measurement resolution Accuracy (1 year) ± (% reading + offset)
±500 mA1 10 uA 0.05% + 100 μA
±100 mA 1 uA 0.05% + 10 μA
±10 mA 100 nA 0.05% + 5 μA
±1 mA 10 nA 0.05% + 500 nA
±100 μA 1nA 0.05% + 50 nA
±10 μA 100 pA 0.05% + 20 nA
Temperature coefficient ±(0.15 × Accuracy)/°C (0℃-18℃,28℃-50℃)

1,500mA range only supports 6V voltage range

I-V output capacity

Output setup time

Output Range Typical output setup time Test conditions:
Fast1,2 Normal1 Slow1
Voltage source 30V <250μS <780μS <2.8ms The time required to reach within 0.1% of the final value under open circuit load. In steps ranging from 10% to 90%.
10V <250μS <780μS <2.8ms
Current source ±500 mA <50 μS <330μS <2.5ms The time required to reach within 0.1% of the final value (0.3% for the 3 a range) under short circuit conditions. Steps are 10% to 90% of the range
±100 mA <50 μS <270μS <2.5ms
±10 mA <50 μS <270μS <2.5ms
±1 mA <50 μS <290μS <2.5ms
±100 μA <50 μS <5mS <2.5ms
±10 μA <150 μS <3mS <2.5m

1,Output conversion rate:Fast,Normal,Slow。

2,The output of fast mode may have large overshoot under different ranges or load conditions. Slow mode is recommended for overshoot sensitive equipment.

Sampling rate and NPLC settings

Collocation method Configuration scope
NPLC 0.0001PLC ~ 10PLC
Sampling Rate 5sps ~ 500Ksps

Derating of measurement accuracy(PLC<1)
Error: percentage of range increase

PLC Range
6V 、30V 10uA 100uA to 100mA 500mA
0.1 0.01% 0.03% 0.01% 0.02
0.01 0.03% 0.06% 0.02% 0.04%
0.001 0.30% 0.40% 0.30% 0.40%

Supplementary features

Sensing mode 2-wire or 4-wire (remote sensing) connection
Maximum sensing lead resistance 1 kΩ(accuracy rating)
Maximum voltage between remote sensing output terminal and sensing terminal 2V
Maximum output voltage of output connector >Full scale 105%
DC floating ground voltage ±60 V DC
SWEEP scanning The scanning interval is configurable from 20us to 16S, and the maximum scanning point is 64K
Auto Range Support. It is recommended to switch off the output before switching the range for overshoot sensitive equipment
Delay measurement(SOURCE DELAY) Support. It is recommended that the user set the appropriate source delay to obtain more accurate measurement values
Over temperature protection When the internal temperature is detected to be too high, the output will be turned off, and the operation will be resumed when the temperature returns below 65 ℃
Other output abnormal protection Power off and restart to recover operation or hardware damage

Environmental indicators

Environment Use in indoor facilities
Work 0 ° C to + 50 ° C, 30% to 70% relative humidity non condensing
Store -30 ° C to 70 ° C, 10% to 90% relative humidity non condensing
Altitude Working height: 0 m to 2000 m, storage: 0 m to 4600 M
Preheat 1 hour